| 48 T | $370.00 |
| 96 T | $528.00 |
| 96 T*5 | $2244.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.036 | 1.932 |
| 500.00 | 1.582 | 1.478 |
| 250.00 | 1.012 | 0.908 |
| 125.00 | 0.739 | 0.635 |
| 62.50 | 0.433 | 0.329 |
| 31.25 | 0.352 | 0.248 |
| 15.63 | 0.235 | 0.131 |
| 0.00 | 0.104 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 94-107% | 100% |
| EDTA plasma(n=5) | 82-95% | 88% |
| Heparin plasma(n=5) | 94-105% | 98% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-100% | 80-97% | 81-92% | 87-103% |
| EDTA plasma(n=5) | 82-93% | 95-104% | 82-96% | 85-98% |
| Heparin plasma(n=5) | 95-104% | 82-96% | 92-105% | 83-92% |