| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (µg/mL) | OD | Corrected OD |
|---|---|---|
| 200.00 | 2.067 | 1.974 |
| 100.00 | 1.646 | 1.553 |
| 50.00 | 1.161 | 1.068 |
| 25.00 | 0.850 | 0.757 |
| 12.50 | 0.513 | 0.420 |
| 6.25 | 0.309 | 0.216 |
| 3.13 | 0.173 | 0.080 |
| 0.00 | 0.093 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 91-104% | 98% |
| EDTA plasma(n=5) | 94-105% | 100% |
| Heparin plasma(n=5) | 89-99% | 94% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 92-105% | 85-101% | 91-116% | 87-95% |
| EDTA plasma(n=5) | 83-110% | 92-100% | 82-103% | 82-109% |
| Heparin plasma(n=5) | 87-99% | 94-107% | 88-112% | 86-108% |