| 48 T | $370.00 |
| 96 T | $528.00 |
| 96 T*5 | $2244.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 500.00 | 1.994 | 1.899 |
| 250.00 | 1.115 | 1.020 |
| 125.00 | 0.752 | 0.657 |
| 62.50 | 0.627 | 0.532 |
| 31.25 | 0.385 | 0.290 |
| 15.63 | 0.295 | 0.200 |
| 7.82 | 0.152 | 0.057 |
| 0.00 | 0.095 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 92-106% | 99% |
| EDTA plasma(n=5) | 88-103% | 95% |
| Heparin plasma(n=5) | 77-95% | 86% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-97% | 87-96% | 86-98% | 91-103% |
| EDTA plasma(n=5) | 87-90% | 87-98% | 85-96% | 85-95% |
| Heparin plasma(n=5) | 87-98% | 85-96% | 92-102% | 81-93% |