| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (µg/mL) | OD | Corrected OD |
|---|---|---|
| 500.00 | 1.952 | 1.880 |
| 250.00 | 1.643 | 1.571 |
| 125.00 | 1.174 | 1.102 |
| 62.50 | 0.963 | 0.891 |
| 31.25 | 0.508 | 0.436 |
| 15.63 | 0.273 | 0.201 |
| 7.82 | 0.135 | 0.063 |
| 0.00 | 0.072 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 78-92% | 85% |
| EDTA plasma(n=5) | 92-105% | 98% |
| Heparin plasma(n=5) | 85-97% | 91% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-105% | 87-98% | 91-102% | 86-102% |
| EDTA plasma(n=5) | 86-94% | 85-95% | 89-99% | 85-95% |
| Heparin plasma(n=5) | 89-97% | 85-94% | 79-96% | 85-99% |