| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (µg/mL) | OD | Corrected OD |
|---|---|---|
| 500.00 | 2.017 | 1.949 |
| 250.00 | 1.456 | 1.388 |
| 125.00 | 0.912 | 0.844 |
| 62.50 | 0.647 | 0.579 |
| 31.25 | 0.438 | 0.370 |
| 15.63 | 0.224 | 0.156 |
| 7.81 | 0.144 | 0.076 |
| 0.00 | 0.068 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 90-107% | 99% |
| EDTA plasma(n=5) | 94-123% | 109% |
| Heparin plasma(n=5) | 84-101% | 93% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 81-98% | 85-119% | 81-118% | 85-98% |
| EDTA plasma(n=5) | 94-117% | 87-116% | 81-111% | 82-103% |
| Heparin plasma(n=5) | 81-88% | 85-92% | 93-105% | 95-103% |