| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.330 | 2.240 |
| 500.00 | 1.689 | 1.599 |
| 250.00 | 0.829 | 0.739 |
| 125.00 | 0.505 | 0.415 |
| 62.50 | 0.345 | 0.255 |
| 31.25 | 0.269 | 0.179 |
| 15.63 | 0.140 | 0.050 |
| 0.00 | 0.090 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 87-100% | 94% |
| EDTA plasma(n=5) | 84-128% | 106% |
| Heparin plasma(n=5) | 88-117% | 103% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 81-106% | 89-111% | 91-106% | 89-95% |
| EDTA plasma(n=5) | 81-86% | 88-95% | 84-98% | 85-93% |
| Heparin plasma(n=5) | 87-120% | 88-113% | 83-106% | 91-118% |