| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 4000.00 | 2.180 | 2.051 |
| 2000.00 | 1.811 | 1.682 |
| 1000.00 | 1.228 | 1.099 |
| 500.00 | 0.883 | 0.754 |
| 250.00 | 0.538 | 0.409 |
| 125.00 | 0.350 | 0.221 |
| 62.50 | 0.254 | 0.125 |
| 0.00 | 0.129 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 93-102% | 98% |
| EDTA plasma(n=5) | 82-102% | 92% |
| Heparin plasma(n=5) | 84-106% | 95% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 89-113% | 94-104% | 89-97% | 94-116% |
| EDTA plasma(n=5) | 92-107% | 92-102% | 86-104% | 80-113% |
| Heparin plasma(n=5) | 88-107% | 86-103% | 91-99% | 89-114% |