| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.336 | 2.245 |
| 500.00 | 1.694 | 1.603 |
| 250.00 | 0.805 | 0.714 |
| 125.00 | 0.497 | 0.406 |
| 62.50 | 0.352 | 0.261 |
| 31.25 | 0.273 | 0.182 |
| 15.63 | 0.142 | 0.051 |
| 0.00 | 0.091 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 81-95% | 88% |
| EDTA plasma(n=5) | 87-99% | 93% |
| Heparin plasma(n=5) | 87-99% | 93% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-94% | 83-96% | 87-98% | 87-100% |
| EDTA plasma(n=5) | 88-99% | 81-93% | 89-97% | 85-94% |
| Heparin plasma(n=5) | 87-96% | 82-95% | 93-102% | 87-96% |