| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 1.873 | 1.776 |
| 500.00 | 1.506 | 1.409 |
| 250.00 | 1.153 | 1.056 |
| 125.00 | 0.890 | 0.793 |
| 62.50 | 0.562 | 0.465 |
| 31.25 | 0.312 | 0.215 |
| 15.63 | 0.228 | 0.131 |
| 0.00 | 0.097 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 88-98% | 93% |
| EDTA plasma(n=5) | 89-102% | 96% |
| Heparin plasma(n=5) | 93-119% | 106% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 93-104% | 86-97% | 87-118% | 81-109% |
| EDTA plasma(n=5) | 89-104% | 82-102% | 90-119% | 92-119% |
| Heparin plasma(n=5) | 89-102% | 83-99% | 81-108% | 94-116% |