| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.176 | 2.105 |
| 5.00 | 1.677 | 1.606 |
| 2.50 | 1.155 | 1.084 |
| 1.25 | 0.723 | 0.652 |
| 0.63 | 0.516 | 0.445 |
| 0.31 | 0.288 | 0.217 |
| 0.16 | 0.175 | 0.104 |
| 0.00 | 0.071 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 93-118% | 106% |
| EDTA plasma(n=5) | 82-119% | 101% |
| Heparin plasma(n=5) | 82-104% | 93% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 81-88% | 90-112% | 90-113% | 88-113% |
| EDTA plasma(n=5) | 89-103% | 86-107% | 84-109% | 86-101% |
| Heparin plasma(n=5) | 88-117% | 87-99% | 88-116% | 91-116% |