| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (µg/mL) | OD | Corrected OD |
|---|---|---|
| 200.00 | 2.201 | 2.114 |
| 100.00 | 1.608 | 1.521 |
| 50.00 | 1.106 | 1.019 |
| 25.00 | 0.933 | 0.846 |
| 12.50 | 0.614 | 0.527 |
| 6.25 | 0.357 | 0.270 |
| 3.13 | 0.226 | 0.139 |
| 0.00 | 0.087 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 80-96% | 88% |
| EDTA plasma(n=5) | 91-105% | 98% |
| Heparin plasma(n=5) | 82-95% | 88% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 95-103% | 95-103% | 84-98% | 85-92% |
| EDTA plasma(n=5) | 83-101% | 87-98% | 79-92% | 86-97% |
| Heparin plasma(n=5) | 83-96% | 93-107% | 88-101% | 90-99% |