48 T | $412.00 |
96 T | $588.00 |
Concentration (µg/mL) | OD | Corrected OD |
---|---|---|
50.00 | 1.948 | 1.859 |
25.00 | 1.720 | 1.631 |
12.50 | 1.225 | 1.136 |
6.25 | 0.871 | 0.782 |
3.13 | 0.518 | 0.429 |
1.57 | 0.319 | 0.230 |
0.79 | 0.226 | 0.137 |
0.00 | 0.089 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 92-107% | 99% |
EDTA plasma(n=5) | 83-96% | 89% |
Heparin plasma(n=5) | 82-98% | 90% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 93-101% | 79-96% | 89-100% | 85-96% |
EDTA plasma(n=5) | 87-99% | 85-96% | 81-95% | 95-105% |
Heparin plasma(n=5) | 88-102% | 93-101% | 90-101% | 87-98% |