| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 20.00 | 2.079 | 1.984 |
| 10.00 | 1.590 | 1.495 |
| 5.00 | 1.286 | 1.191 |
| 2.50 | 0.897 | 0.802 |
| 1.25 | 0.543 | 0.448 |
| 0.63 | 0.342 | 0.247 |
| 0.32 | 0.235 | 0.140 |
| 0.00 | 0.095 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 78-92% | 85% |
| EDTA plasma(n=5) | 84-96% | 90% |
| Heparin plasma(n=5) | 87-99% | 93% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 88-101% | 90-99% | 81-97% | 82-96% |
| EDTA plasma(n=5) | 95-102% | 85-92% | 90-96% | 87-1028% |
| Heparin plasma(n=5) | 85-97% | 96-105% | 87-96% | 93-108% |