| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 20.00 | 0.199 | |
| 10.00 | 0.467 | |
| 5.00 | 0.589 | |
| 2.50 | 0.927 | |
| 1.25 | 1.275 | |
| 0.63 | 1.596 | |
| 0.32 | 1.869 | |
| 0.00 | 2.219 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 87-95% | 91% |
| EDTA plasma(n=5) | 82-95% | 88% |
| Heparin plasma(n=5) | 78-96% | 87% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 95-103% | 89-101% | 87-98% | 85-92% |
| EDTA plasma(n=5) | 95-102% | 87-98% | 86-97% | 83-96% |
| Heparin plasma(n=5) | 85-94% | 97-103% | 79-91% | 82-91% |