| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 2000.00 | 0.175 | |
| 1000.00 | 0.386 | |
| 500.00 | 0.689 | |
| 250.00 | 0.895 | |
| 125.00 | 1.389 | |
| 62.50 | 1.565 | |
| 31.25 | 1.861 | |
| 0.00 | 2.132 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 81-95% | 88% |
| EDTA plasma(n=5) | 80-95% | 87% |
| Heparin plasma(n=5) | 87-99% | 91% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 81-96% | 87-98% | 95-101% | 87-94% |
| EDTA plasma(n=5) | 95-102% | 83-96% | 97-103% | 85-92% |
| Heparin plasma(n=5) | 88-104% | 96-105% | 92-103% | 82-96% |