| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (µg/mL) | OD | Corrected OD |
|---|---|---|
| 100.00 | 2.185 | 2.090 |
| 50.00 | 1.502 | 1.407 |
| 25.00 | 1.174 | 1.079 |
| 12.50 | 0.875 | 0.780 |
| 6.25 | 0.614 | 0.519 |
| 3.13 | 0.354 | 0.259 |
| 1.57 | 0.179 | 0.084 |
| 0.00 | 0.095 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 88-99% | 93% |
| EDTA plasma(n=5) | 80-97% | 88% |
| Heparin plasma(n=5) | 82-95% | 88% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 79-92% | 95-103% | 82-98% | 85-97% |
| EDTA plasma(n=5) | 86-96% | 93-104% | 89-97% | 85-94% |
| Heparin plasma(n=5) | 97-105% | 88-95% | 96-102% | 83-97% |