| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.058 | 1.976 |
| 5.00 | 1.577 | 1.495 |
| 2.50 | 1.032 | 0.950 |
| 1.25 | 0.795 | 0.713 |
| 0.63 | 0.562 | 0.480 |
| 0.32 | 0.357 | 0.275 |
| 0.16 | 0.196 | 0.114 |
| 0.00 | 0.082 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 78-92% | 85% |
| EDTA plasma(n=5) | 88-106% | 97% |
| Heparin plasma(n=5) | 81-95% | 88% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 78-92% | 82-96% | 86-99% | 93-101% |
| EDTA plasma(n=5) | 85-96% | 86-98% | 92-103% | 87-93% |
| Heparin plasma(n=5) | 81-94% | 85-97% | 93-102% | 81-95% |