| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 100.00 | 1.989 | 1.898 |
| 50.00 | 1.562 | 1.471 |
| 25.00 | 1.139 | 1.048 |
| 12.50 | 0.897 | 0.806 |
| 6.25 | 0.483 | 0.392 |
| 3.13 | 0.426 | 0.335 |
| 1.57 | 0.190 | 0.099 |
| 0.00 | 0.091 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 87-99% | 93% |
| EDTA plasma(n=5) | 80-93% | 86% |
| Heparin plasma(n=5) | 93-107% | 100% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 89-101% | 87-98% | 92-104% | 82-96% |
| EDTA plasma(n=5) | 88-99% | 81-93% | 89-97% | 85-94% |
| Heparin plasma(n=5) | 95-104% | 93-105% | 82-101% | 91-103% |