| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (µg/mL) | OD | Corrected OD |
|---|---|---|
| 100.00 | 0.191 | |
| 50.00 | 0.352 | |
| 25.00 | 0.567 | |
| 12.50 | 0.861 | |
| 6.25 | 1.112 | |
| 3.13 | 1.585 | |
| 1.57 | 1.697 | |
| 0.00 | 2.211 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 92-106% | 99% |
| EDTA plasma(n=5) | 88-103% | 95% |
| Heparin plasma(n=5) | 77-95% | 91% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-98% | 97-104% | 95-104% | 82-90% |
| EDTA plasma(n=5) | 85-93% | 87-96% | 97-105% | 95-102% |
| Heparin plasma(n=5) | 89-97% | 81-92% | 78-87% | 91-99% |