| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.207 | 2.111 |
| 500.00 | 1.587 | 1.491 |
| 250.00 | 1.118 | 1.022 |
| 125.00 | 0.861 | 0.765 |
| 62.50 | 0.559 | 0.463 |
| 31.25 | 0.265 | 0.169 |
| 15.63 | 0.325 | 0.229 |
| 0.00 | 0.096 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 91-105% | 98% |
| EDTA plasma(n=5) | 85-97% | 91% |
| Heparin plasma(n=5) | 80-94% | 87% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 88-103% | 83-94% | 92-105% | 89-97% |
| EDTA plasma(n=5) | 85-92% | 89-96% | 90-98% | 87-98% |
| Heparin plasma(n=5) | 92-104% | 86-98% | 80-93% | 89-97% |