| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 2000.00 | 0.175 | |
| 1000.00 | 0.479 | |
| 500.00 | 0.675 | |
| 250.00 | 0.959 | |
| 125.00 | 1.156 | |
| 62.50 | 1.511 | |
| 31.25 | 1.755 | |
| 0.00 | 2.226 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 87-99% | 93% |
| EDTA plasma(n=5) | 90-103% | 96% |
| Heparin plasma(n=5) | 78-93% | 98% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 81-94% | 96-105% | 93-106% | 82-95% |
| EDTA plasma(n=5) | 89-100% | 79-91% | 89-103% | 82-95% |
| Heparin plasma(n=5) | 87-98% | 91-105% | 87-98% | 89-94% |