| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 1.989 | 1.882 |
| 500.00 | 1.567 | 1.460 |
| 250.00 | 1.198 | 1.091 |
| 125.00 | 0.933 | 0.826 |
| 62.50 | 0.651 | 0.544 |
| 31.25 | 0.335 | 0.228 |
| 15.63 | 0.262 | 0.155 |
| 0.00 | 0.107 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 87-99% | 93% |
| EDTA plasma(n=5) | 90-103% | 96% |
| Heparin plasma(n=5) | 78-93% | 85% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 93-105% | 88-102% | 87-98% | 85-96% |
| EDTA plasma(n=5) | 97-105% | 86-92% | 95-103% | 92-104% |
| Heparin plasma(n=5) | 85-94% | 92-101% | 87-96% | 82-90% |