| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 1.941 | 1.847 |
| 500.00 | 1.657 | 1.563 |
| 250.00 | 1.157 | 1.063 |
| 125.00 | 0.845 | 0.751 |
| 62.50 | 0.411 | 0.317 |
| 31.25 | 0.352 | 0.258 |
| 15.63 | 0.247 | 0.153 |
| 0.00 | 0.094 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 83-95% | 89% |
| EDTA plasma(n=5) | 77-95% | 86% |
| Heparin plasma(n=5) | 85-97% | 91% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-94% | 95-103% | 79-97% | 93-102% |
| EDTA plasma(n=5) | 95-102% | 93-101% | 97-105% | 95-103% |
| Heparin plasma(n=5) | 93-101% | 86-92% | 81-93% | 79-95% |