| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 2000.00 | 0.175 | |
| 1000.00 | 0.396 | |
| 500.00 | 0.715 | |
| 250.00 | 0.812 | |
| 125.00 | 1.275 | |
| 62.50 | 1.572 | |
| 31.25 | 1.624 | |
| 0.00 | 2.232 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 78-90% | 84% |
| EDTA plasma(n=5) | 84-97% | 90% |
| Heparin plasma(n=5) | 90-105% | 92% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 93-102% | 81-96% | 79-94% | 87-98% |
| EDTA plasma(n=5) | 86-97% | 93-102% | 87-98% | 83-97% |
| Heparin plasma(n=5) | 86-94% | 93-101% | 86-97% | 90-101% |