| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 1.948 | 1.862 |
| 5.00 | 1.612 | 1.526 |
| 2.50 | 1.285 | 1.199 |
| 1.25 | 0.912 | 0.826 |
| 0.63 | 0.627 | 0.541 |
| 0.32 | 0.315 | 0.229 |
| 0.16 | 0.158 | 0.072 |
| 0.00 | 0.086 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 95-107% | 101% |
| EDTA plasma(n=5) | 90-105% | 97% |
| Heparin plasma(n=5) | 82-95% | 88% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 81-94% | 93-107% | 93-102% | 91-105% |
| EDTA plasma(n=5) | 89-102% | 82-94% | 85-96% | 78-90% |
| Heparin plasma(n=5) | 92-106% | 93-105% | 85-99% | 90-99% |