Products

Horse E2(Estradiol) ELISA Kit
One-click to copy product information
ELK10910
Size: Price:
48 T $265.00
96 T $376.00

Overview

Product name: Horse E2(Estradiol) ELISA Kit
Reactivity: Horse
Alternative Names: 17B-Estradiol; Oestradiol; Beta-Estradiol
Assay Type: Competitive Inhibition
Sensitivity: 13.23 pg/mL
Standard: 3000 pg/mL
Detection Range: 46.88-3000 pg/mL
Sample type: serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
Assay length: 2h
Research Area: Endocrinology;Reproductive science;Hormone metabolism;
Test principle: This assay employs the competitive inhibition enzyme immunoassay technique. The microtiter plate provided in this kit has been pre-coated with Horse E2. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Horse E2. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Horse E2 in the samples is then determined by comparing the OD of the samples to the standard curve.

Standard curve

Concentration (pg/mL) OD Corrected OD
3000.00 0.227
1500.00 0.375
750.00 0.583
375.00 0.897
187.50 1.197
93.75 1.511
46.88 1.813
0.00 2.211

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant E2 and the recovery rates were calculated by comparing the measured value to the expected amount of E2 in samples.
Matrix Recovery range Average
serum(n=5) 93-101% 97%
EDTA plasma(n=5) 85-103% 94%
Heparin plasma(n=5) 95-105% 88%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of E2 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 92-101% 82-98% 85-96% 85-95%
EDTA plasma(n=5) 86-99% 93-107% 85-94% 87-101%
Heparin plasma(n=5) 92-101% 89-97% 85-94% 87-101%