| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 0.191 | |
| 500.00 | 0.427 | |
| 250.00 | 0.596 | |
| 125.00 | 0.933 | |
| 62.50 | 1.197 | |
| 31.25 | 1.511 | |
| 15.63 | 1.761 | |
| 0.00 | 2.123 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 78-95% | 86% |
| EDTA plasma(n=5) | 80-96% | 88% |
| Heparin plasma(n=5) | 80-92% | 85% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 87-96% | 85-97% | 86-98% | 91-103% |
| EDTA plasma(n=5) | 85-104% | 83-96% | 95-103% | 88-95% |
| Heparin plasma(n=5) | 80-101% | 87-101% | 86-97% | 88-95% |