| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 2000.00 | 0.197 | |
| 1000.00 | 0.375 | |
| 500.00 | 0.611 | |
| 250.00 | 0.873 | |
| 125.00 | 1.197 | |
| 62.50 | 1.583 | |
| 31.25 | 1.813 | |
| 0.00 | 2.211 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 79-93% | 86% |
| EDTA plasma(n=5) | 87-93% | 90% |
| Heparin plasma(n=5) | 81-95% | 91% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 88-97% | 96-105% | 92-103% | 81-90% |
| EDTA plasma(n=5) | 85-97% | 83-96% | 95-103% | 88-95% |
| Heparin plasma(n=5) | 87-98% | 85-94% | 87-101% | 89-99% |