| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 0.199 | |
| 500.00 | 0.397 | |
| 250.00 | 0.611 | |
| 125.00 | 0.823 | |
| 62.50 | 1.112 | |
| 31.25 | 1.583 | |
| 15.63 | 1.697 | |
| 0.00 | 2.213 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 80-95% | 87% |
| EDTA plasma(n=5) | 86-99% | 92% |
| Heparin plasma(n=5) | 83-95% | 88% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 81-97% | 88-102% | 78-90% | 91-99% |
| EDTA plasma(n=5) | 87-102% | 85-94% | 87-101% | 89-99% |
| Heparin plasma(n=5) | 88-104% | 96-105% | 92-103% | 82-96% |