| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 2000.00 | 1.978 | 1.893 |
| 1000.00 | 1.593 | 1.508 |
| 500.00 | 1.047 | 0.962 |
| 250.00 | 0.949 | 0.864 |
| 125.00 | 0.572 | 0.487 |
| 62.50 | 0.355 | 0.270 |
| 31.25 | 0.175 | 0.090 |
| 0.00 | 0.085 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 87-100% | 94% |
| EDTA plasma(n=5) | 92-102% | 97% |
| Heparin plasma(n=5) | 93-99% | 96% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 86-101% | 88-102% | 84-103% | 93-105% |
| EDTA plasma(n=5) | 80-115% | 84-114% | 86-118% | 84-108% |
| Heparin plasma(n=5) | 91-107% | 87-105% | 87-119% | 83-104% |