| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 2000.00 | 1.994 | 1.907 |
| 1000.00 | 1.587 | 1.500 |
| 500.00 | 1.035 | 0.948 |
| 250.00 | 0.933 | 0.846 |
| 125.00 | 0.556 | 0.469 |
| 62.50 | 0.351 | 0.264 |
| 31.25 | 0.175 | 0.088 |
| 0.00 | 0.087 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 79-97% | 88% |
| EDTA plasma(n=5) | 80-95% | 87% |
| Heparin plasma(n=5) | 84-98% | 91% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-92% | 90-96% | 79-102% | 83-99% |
| EDTA plasma(n=5) | 81-94% | 96-105% | 92-103% | 82-96% |
| Heparin plasma(n=5) | 78-92% | 83-96% | 95-107% | 87-102% |