| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 100.00 | 1.908 | 1.840 |
| 50.00 | 1.574 | 1.506 |
| 25.00 | 1.183 | 1.115 |
| 12.50 | 0.839 | 0.771 |
| 6.25 | 0.589 | 0.521 |
| 3.13 | 0.317 | 0.249 |
| 1.57 | 0.225 | 0.157 |
| 0.00 | 0.068 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 90-107% | 98% |
| EDTA plasma(n=5) | 87-99% | 93% |
| Heparin plasma(n=5) | 82-98% | 90% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 89-102% | 79-94% | 83-96% | 85-106% |
| EDTA plasma(n=5) | 85-107% | 84-98% | 95-103% | 86-98% |
| Heparin plasma(n=5) | 86-93% | 91-105% | 87-98% | 99-102% |