| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 1.916 | 1.830 |
| 5.00 | 1.647 | 1.561 |
| 2.50 | 1.128 | 1.042 |
| 1.25 | 0.877 | 0.791 |
| 0.63 | 0.457 | 0.371 |
| 0.32 | 0.324 | 0.238 |
| 0.16 | 0.206 | 0.120 |
| 0.00 | 0.086 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 92-107% | 99% |
| EDTA plasma(n=5) | 93-107% | 100% |
| Heparin plasma(n=5) | 87-99% | 93% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-101% | 78-96% | 79-98% | 93-102% |
| EDTA plasma(n=5) | 87-96% | 82-95% | 93-102% | 87-96% |
| Heparin plasma(n=5) | 85-94% | 83-96% | 87-98% | 87-100% |