Products

Cattle Cys-C(Cystatin C) ELISA Kit
One-click to copy product information
ELK10838
Size: Price:
48 T $412.00
96 T $588.00

Overview

Product name: Cattle Cys-C(Cystatin C) ELISA Kit
Reactivity: Cattle
Alternative Names: CST3; Cystatin 3; ARMD11; Post-Gamma-Globulin; Gamma Trace; Post-Gamma-Globulin; Neuroendocrine Basic Polypeptide; Amyloid Angiopathy And Cerebral Hemorrhage
Assay Type: Sandwich
Sensitivity: 0.47 ng/mL
Standard: 100 ng/mL
Detection Range: 1.57-100 ng/mL
Sample type: serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
Assay length: 3.5h
Research Area: Kidney biomarker;Urology;
Uniprot ID: P01035
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Cattle Cys-C. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Cattle Cys-C. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Cattle Cys-C, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Cattle Cys-C in the samples is then determined by comparing the OD of the samples to the standard curve.

Standard curve

Concentration (ng/mL) OD Corrected OD
100.00 2.056 1.971
50.00 1.594 1.509
25.00 1.142 1.057
12.50 0.806 0.721
6.25 0.453 0.368
3.13 0.308 0.223
1.57 0.135 0.050
0.00 0.085 0.000

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant Cys-C and the recovery rates were calculated by comparing the measured value to the expected amount of Cys-C in samples.
Matrix Recovery range Average
serum(n=5) 78-90% 84%
EDTA plasma(n=5) 95-107% 101%
Heparin plasma(n=5) 90-105% 97%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Cys-C and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 89-97% 85-101% 84-98% 85-93%
EDTA plasma(n=5) 93-104% 87-96% 82-95% 78-99%
Heparin plasma(n=5) 79-92% 89-102% 87-98% 80-92%