| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (µg/mL) | OD | Corrected OD |
|---|---|---|
| 500.00 | 2.268 | 2.183 |
| 250.00 | 1.527 | 1.442 |
| 125.00 | 1.263 | 1.178 |
| 62.50 | 0.869 | 0.784 |
| 31.25 | 0.614 | 0.529 |
| 15.63 | 0.332 | 0.247 |
| 7.82 | 0.291 | 0.206 |
| 0.00 | 0.085 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 95-107% | 101% |
| EDTA plasma(n=5) | 82-95% | 88% |
| Heparin plasma(n=5) | 80-94% | 87% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-96% | 87-101% | 89-102% | 85-98% |
| EDTA plasma(n=5) | 85-94% | 79-96% | 92-101% | 83-90% |
| Heparin plasma(n=5) | 95-102% | 93-101% | 97-105% | 95-103% |