| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 500.00 | 2.056 | 1.960 |
| 250.00 | 1.593 | 1.497 |
| 125.00 | 1.196 | 1.100 |
| 62.50 | 0.835 | 0.739 |
| 31.25 | 0.511 | 0.415 |
| 15.63 | 0.356 | 0.260 |
| 7.82 | 0.228 | 0.132 |
| 0.00 | 0.096 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 83-97% | 90% |
| EDTA plasma(n=5) | 80-97% | 88% |
| Heparin plasma(n=5) | 85-101% | 93% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-94% | 87-96% | 92-105% | 89-102% |
| EDTA plasma(n=5) | 85-98% | 97-104% | 92-101% | 82-96% |
| Heparin plasma(n=5) | 95-104% | 97-106% | 85-98% | 81-92% |