| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 50.00 | 2.058 | 1.954 |
| 25.00 | 1.472 | 1.368 |
| 12.50 | 1.183 | 1.079 |
| 6.25 | 0.752 | 0.648 |
| 3.13 | 0.545 | 0.441 |
| 1.57 | 0.369 | 0.265 |
| 0.79 | 0.216 | 0.112 |
| 0.00 | 0.104 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 92-105% | 98% |
| EDTA plasma(n=5) | 85-97% | 91% |
| Heparin plasma(n=5) | 78-92% | 85% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 92-101% | 87-98% | 85-92% | 79-96% |
| EDTA plasma(n=5) | 88-97% | 90-99% | 87-103% | 82-97% |
| Heparin plasma(n=5) | 85-94% | 90-99% | 82-101% | 91-102% |