| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 400.00 | 2.045 | 1.948 |
| 200.00 | 1.528 | 1.431 |
| 100.00 | 1.185 | 1.088 |
| 50.00 | 0.875 | 0.778 |
| 25.00 | 0.557 | 0.460 |
| 12.50 | 0.325 | 0.228 |
| 6.25 | 0.213 | 0.116 |
| 0.00 | 0.097 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 84-106% | 100% |
| EDTA plasma(n=5) | 87-93% | 90% |
| Heparin plasma(n=5) | 80-90% | 85% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 93-104% | 87-96% | 82-95% | 78-99% |
| EDTA plasma(n=5) | 95-104% | 93-105% | 82-101% | 91-103% |
| Heparin plasma(n=5) | 89-97% | 85-101% | 84-98% | 85-93% |