| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (µg/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 1.941 | 1.847 |
| 5.00 | 1.609 | 1.515 |
| 2.50 | 1.106 | 1.012 |
| 1.25 | 0.865 | 0.771 |
| 0.63 | 0.545 | 0.451 |
| 0.32 | 0.357 | 0.263 |
| 0.16 | 0.175 | 0.081 |
| 0.00 | 0.094 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 80-93% | 86% |
| EDTA plasma(n=5) | 80-95% | 87% |
| Heparin plasma(n=5) | 95-107% | 101% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 79-91% | 87-96% | 83-101% | 82-95% |
| EDTA plasma(n=5) | 87-102% | 85-94% | 87-101% | 89-99% |
| Heparin plasma(n=5) | 83-97% | 86-97% | 90-98% | 80-93% |