| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (µg/mL) | OD | Corrected OD |
|---|---|---|
| 100.00 | 2.154 | 2.072 |
| 50.00 | 1.557 | 1.475 |
| 25.00 | 1.147 | 1.065 |
| 12.50 | 0.895 | 0.813 |
| 6.25 | 0.577 | 0.495 |
| 3.13 | 0.315 | 0.233 |
| 1.57 | 0.226 | 0.144 |
| 0.00 | 0.082 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 83-99% | 95% |
| EDTA plasma(n=5) | 85-97% | 91% |
| Heparin plasma(n=5) | 85-97% | 91% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-94% | 97-106% | 89-97% | 95-104% |
| EDTA plasma(n=5) | 92-103% | 83-98% | 88-97% | 96-104% |
| Heparin plasma(n=5) | 85-97% | 84-100% | 95-103% | 88-104% |