| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 100.00 | 1.946 | 1.843 |
| 50.00 | 1.563 | 1.460 |
| 25.00 | 1.088 | 0.985 |
| 12.50 | 0.847 | 0.744 |
| 6.25 | 0.539 | 0.436 |
| 3.13 | 0.397 | 0.294 |
| 1.57 | 0.192 | 0.089 |
| 0.00 | 0.103 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 95-107% | 101% |
| EDTA plasma(n=5) | 85-97% | 91% |
| Heparin plasma(n=5) | 91-105% | 98% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 88-97% | 86-92% | 91-99% | 79-97% |
| EDTA plasma(n=5) | 87-96% | 85-97% | 85-94% | 87-96% |
| Heparin plasma(n=5) | 85-97% | 87-96% | 85-96% | 85-95% |