| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 20.00 | 2.136 | 2.033 |
| 10.00 | 1.720 | 1.617 |
| 5.00 | 1.224 | 1.121 |
| 2.50 | 0.852 | 0.749 |
| 1.25 | 0.527 | 0.424 |
| 0.63 | 0.325 | 0.222 |
| 0.32 | 0.189 | 0.086 |
| 0.00 | 0.103 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 95-104% | 98% |
| EDTA plasma(n=5) | 82-90% | 86% |
| Heparin plasma(n=5) | 84-92% | 87% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 89-97% | 85-94% | 79-96% | 85-99% |
| EDTA plasma(n=5) | 83-95% | 88-102% | 85-97% | 80-92% |
| Heparin plasma(n=5) | 98-107% | 93-102% | 80-93% | 91-100% |