| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 200.00 | 2.168 | 2.073 |
| 100.00 | 1.628 | 1.533 |
| 50.00 | 1.223 | 1.128 |
| 25.00 | 0.833 | 0.738 |
| 12.50 | 0.545 | 0.450 |
| 6.25 | 0.357 | 0.262 |
| 3.13 | 0.252 | 0.157 |
| 0.00 | 0.095 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 82-94% | 88% |
| EDTA plasma(n=5) | 82-98% | 90% |
| Heparin plasma(n=5) | 80-97% | 88% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 88-98% | 87-103% | 89-97% | 85-94% |
| EDTA plasma(n=5) | 86-96% | 91-107% | 87-101% | 95-102% |
| Heparin plasma(n=5) | 92-103% | 83-98% | 88-97% | 96-104% |