| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (µg/mL) | OD | Corrected OD |
|---|---|---|
| 500.00 | 1.916 | 1.828 |
| 250.00 | 1.593 | 1.505 |
| 125.00 | 1.174 | 1.086 |
| 62.50 | 0.852 | 0.764 |
| 31.25 | 0.563 | 0.475 |
| 15.63 | 0.366 | 0.278 |
| 7.82 | 0.175 | 0.087 |
| 0.00 | 0.088 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 84-97% | 90% |
| EDTA plasma(n=5) | 90-105% | 97% |
| Heparin plasma(n=5) | 78-90% | 84% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 97-105% | 86-98% | 84-93% | 89-102% |
| EDTA plasma(n=5) | 89-102% | 92-101% | 88-102% | 87-102% |
| Heparin plasma(n=5) | 92-101% | 87-98% | 85-92% | 79-96% |