| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 5000.00 | 2.164 | 2.083 |
| 2500.00 | 1.699 | 1.618 |
| 1250.00 | 1.225 | 1.144 |
| 625.00 | 0.864 | 0.783 |
| 312.50 | 0.593 | 0.512 |
| 156.25 | 0.402 | 0.321 |
| 78.13 | 0.252 | 0.171 |
| 0.00 | 0.081 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 93-107% | 100% |
| EDTA plasma(n=5) | 87-99% | 93% |
| Heparin plasma(n=5) | 87-95% | 91% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 88-97% | 79-92% | 81-97% | 87-95% |
| EDTA plasma(n=5) | 85-102% | 83-96% | 87-98% | 82-95% |
| Heparin plasma(n=5) | 98-106% | 86-99% | 78-95% | 89-97% |