| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (µg/mL) | OD | Corrected OD |
|---|---|---|
| 50.00 | 2.126 | 2.042 |
| 25.00 | 1.328 | 1.244 |
| 12.50 | 0.993 | 0.909 |
| 6.25 | 0.791 | 0.707 |
| 3.13 | 0.538 | 0.454 |
| 1.57 | 0.297 | 0.213 |
| 0.79 | 0.235 | 0.151 |
| 0.00 | 0.084 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 88-102% | 95% |
| EDTA plasma(n=5) | 95-108% | 101% |
| Heparin plasma(n=5) | 83-96% | 89% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-97% | 86-102% | 87-99% | 99-105% |
| EDTA plasma(n=5) | 88-96% | 90-101% | 93-102% | 85-94% |
| Heparin plasma(n=5) | 92-103% | 81-92% | 78-87% | 91-99% |