| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 100.00 | 1.953 | 1.865 |
| 50.00 | 1.546 | 1.458 |
| 25.00 | 1.178 | 1.090 |
| 12.50 | 0.820 | 0.732 |
| 6.25 | 0.544 | 0.456 |
| 3.13 | 0.352 | 0.264 |
| 1.56 | 0.259 | 0.171 |
| 0.00 | 0.088 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 90-103% | 97% |
| EDTA plasma(n=5) | 90-106% | 98% |
| Heparin plasma(n=5) | 88-109% | 99% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 81-98% | 82-102% | 93-105% | 81-120% |
| EDTA plasma(n=5) | 86-97% | 90-113% | 82-93% | 91-109% |
| Heparin plasma(n=5) | 95-106% | 86-92% | 80-112% | 90-96% |