| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (µg/mL) | OD | Corrected OD |
|---|---|---|
| 500.00 | 2.154 | 2.071 |
| 250.00 | 1.631 | 1.548 |
| 125.00 | 1.243 | 1.160 |
| 62.50 | 0.933 | 0.850 |
| 31.25 | 0.581 | 0.498 |
| 15.63 | 0.342 | 0.259 |
| 7.82 | 0.196 | 0.113 |
| 0.00 | 0.083 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 80-95% | 87% |
| EDTA plasma(n=5) | 82-97% | 89% |
| Heparin plasma(n=5) | 95-108% | 101% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 95-103% | 87-98% | 85-92% | 79-96% |
| EDTA plasma(n=5) | 85-94% | 87-103% | 81-95% | 82-90% |
| Heparin plasma(n=5) | 85-92% | 90-96% | 79-102% | 83-99% |