| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 1.985 | 1.889 |
| 5.00 | 1.579 | 1.483 |
| 2.50 | 1.114 | 1.018 |
| 1.25 | 0.891 | 0.795 |
| 0.63 | 0.449 | 0.353 |
| 0.31 | 0.302 | 0.206 |
| 0.16 | 0.215 | 0.119 |
| 0.00 | 0.096 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 84-100% | 92% |
| EDTA plasma(n=5) | 88-104% | 96% |
| Heparin plasma(n=5) | 92-119% | 106% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 84-102% | 87-95% | 94-101% | 81-105% |
| EDTA plasma(n=5) | 95-102% | 86-100% | 88-106% | 82-103% |
| Heparin plasma(n=5) | 81-117% | 87-116% | 95-115% | 83-117% |