| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 5000.00 | 2.204 | 2.101 |
| 2500.00 | 1.502 | 1.399 |
| 1250.00 | 1.121 | 1.018 |
| 625.00 | 0.872 | 0.769 |
| 312.50 | 0.457 | 0.354 |
| 156.25 | 0.300 | 0.197 |
| 78.13 | 0.226 | 0.123 |
| 0.00 | 0.103 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 88-96% | 92% |
| EDTA plasma(n=5) | 95-107% | 101% |
| Heparin plasma(n=5) | 81-95% | 88% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 86-93% | 93-101% | 85-98% | 90-101% |
| EDTA plasma(n=5) | 98-105% | 92-101% | 85-105% | 87-98% |
| Heparin plasma(n=5) | 85-92% | 79-96% | 89-100% | 85-96% |